• Spectraval 1501

    Spectraval 1501 is a precise and economic reference grade VIS spectroradiometer for various applications, including monitor and projector calibration and the measurement of different display types. It measures spectral radiance in the visible range and provides a variety of integral data. It has a laser marking of measuring circle during measurement.

    Special versions of the device are spectraval 1501-HiRes (with an increased optical resolution), spectraval 1501-NIR (with a wavelength range up to 1000 nm) and spectraval 1501-LAN (with LAN interface instead of Bluetooth).

    Spectraval 1511

    Spectraval 1511 is a modern spectroradiometer for the VIS wavelength range, which can be used in stand alone mode or with a PC or laptop. It is equipped with a touchpad display and a SD card slot for in field measurements. The available interfaces for operation with the PC software JETI LiVal are USB and Bluetooth.
    Special versions of the device are spectraval 1511-HiRes (with an increased optical resolution), spectraval 1511-NIR (with a wavelength range up to 1000 nm).

    Specbos 1211

    Specbos 1211-2 is a miniaturized and fast broadband spectroradiometer covering the wavelength range from 350-1000 nm (the special version specbos 1211-UV-2 extends the range in UV down to 230 nm). It can be used for various applications both in lab as well as in production and can measure in Radiance mode based on its NIST traceable calibration. It has a laser marking of measuring circle during measurement.
    To extend the device to Irradiance mode, please order the standard diffuser for specbos.
    On special request it is possible to calibrate specbos 1211-2 in Radiant Intensity mode.

  • Model

    Spectraval 1501

    Spectraval 1511

    Specbos 1211

    Spectral range

    380-780nm

    380-1000nm (1501-NIR)

    380-780nm

    380-1000nm (1511-NIR)

    350-1000nm

    230-1000nm (1211-UV)

    Optical bandwidth

    4.5nm

    2nm  (1501-HiRes)

    4.5nm

    2nm (1511-HiRes)

    4.5nm

    Wavelength resolution

    1nm

    DA resolution

    16 bit ADC

    15 bit ADC

    Viewing angle

    1.8°, 20 cm-φ6 mm, 100cm-φ33mm

    Measuring values

    Spectral Radiance/ Irradiance, Luminance/ Illuminance;

    total Radiance/ Irradiance, x,y, u’,v’, CCT, color purity, CRI

    Luminance measuring range

    0.2-180000 cd/m2 (Ill. A)

    0.2-140000 cd/ m2 (Typical warm white LED)

    0.2-150000 cd/m2 (Ill. A)

    0.2-100000 cd/m2 (Typical warm white LED)

    Illuminance measuring range

    -

    1-1800000lx (Ill. A)

    1-1500000lx  (Typical warm white LED)

    Luminance uncertainty

    ±4.4 % (@ 100 cd/m2 and Ill. A):

    Luminance repeatability

    ±1 % (Ill.A)

    Chromaticity uncertainty

    ±0.002 x, y (Ill.A)

    Chromaticity accuracy

    ±0.0005 x, y (Ill.A)

    CCT repeatability

    ±20 K (Ill.A)

    Wavelength accuracy

    0.2nm (HgAr line source)

    0.5nm (HgAr line source)

    Polarization error f8

    <2%

    <1%

    Interface

    USB 2.0, Bluetooth, LAN

    Dimensions

    140 mm×80 mm×70 mm

    140 mm×115 mm×70 mm

    180 mm×82 mm×53 mm

    Weight

    400g

    500g

    450g

    Operation condition

    Temperature 10 -40 °C, humidity < 85 %

    Caliration

    NIST traceable

    Recommended interval

    1 year

  • Display color and white balance correction
    Fluorescent lamp, ultraviolet lamp, light source test
    Testing OLEDs, LEDs, and various flat panel displays

    Plant Lighting PAR Test
    Photobiosafety evaluation testing
    Binning of LEDs